The International journal of oral & maxillofacial implants 2019-Nov/Dec
Effects of Mechanical Instrumentation with Commercially Available Instruments Used in Supportive Peri-implant Therapy: An In Vitro Study.
Само регистрирани потребители могат да превеждат статии
Вход / Регистрация
Линкът е запазен в клипборда
Ключови думи
Резюме
MATERIALS AND METHODS
The coronal one-third of seven RG and seven MA implants was coated with a mixture of cyanoacrylate and toluidine blue dye to resemble calculus. Implants were cleaned with three curettes (SS: stainless steel, PT: plastic, TI: titanium), two ultrasonic tips (UM: metal tip, UP: plastic tip), a titanium brush (TB), and an air-polishing device (AA) until visibly clean. Additionally, a simulation of 1- and 5-year supportive peri-implant therapy (SPT) was performed on 14 implants using the aforementioned instruments with 20 strokes/40 s (T1) or 100 strokes/200 s (T5). Each implant was evaluated using stereomicroscopy, atomic force microscopy, and scanning electron microscopy.RESULTS
UM was the most effective instrument, with 0% average percentage of residual artificial calculus (RAC), followed by TB (2.89%) and UP (4.90%). SS was more effective than TI (15.43% vs 20.12% RAC, respectively), while PT failed to remove any deposit (100% RAC). AA completely removed deposits on RG surfaces but not MA surfaces (26.61% RAC). Noticeable topographic changes were observed between both implant surfaces. RG surfaces became less rough, whereas MA surfaces became rougher at both T1 and T5 with the exception of AA. Plastic- and titanium-like remnants were noted after debridement with PT, SS, and TI, respectively.