Български
Albanian
Arabic
Armenian
Azerbaijani
Belarusian
Bengali
Bosnian
Catalan
Czech
Danish
Deutsch
Dutch
English
Estonian
Finnish
Français
Greek
Haitian Creole
Hebrew
Hindi
Hungarian
Icelandic
Indonesian
Irish
Italian
Japanese
Korean
Latvian
Lithuanian
Macedonian
Mongolian
Norwegian
Persian
Polish
Portuguese
Romanian
Russian
Serbian
Slovak
Slovenian
Spanish
Swahili
Swedish
Turkish
Ukrainian
Vietnamese
Български
中文(简体)
中文(繁體)
Journal of Heart Valve Disease 2003-Jan

Kinematics of synthetic flexible leaflet heart valves during accelerated testing.

Само регистрирани потребители могат да превеждат статии
Вход / Регистрация
Линкът е запазен в клипборда
Stephen S D'Souza
Maya Butterfield
John Fisher

Ключови думи

Резюме

OBJECTIVE

The study aim was to examine the effect of variation of test frequencies on the duration and kinematics of valve leaflets during opening, and at the fully open position.

METHODS

Trileaflet polyurethane heart valves were tested in a Rowan Ash fatigue accelerator at 5, 10 and 15 Hz. Images of the valves were acquired using a high-speed camera and assessed qualitatively during opening and in the fully open position. The bending strains along the leaflet free edge and leaflet belly were also calculated from these images. Results were compared with those found at 1 Hz in a pulsatile flow simulator. The time duration of each phase of the cycle and the strain time integral when opening were also determined.

RESULTS

In the fully open position, the maximum bending strain found at the commissural regions of the valve was not dependent on the cycle rate. During opening, one leaflet always deformed more than either of the others, but this leaflet was not necessarily the same one each time. The opening characteristics of the valves were dependent on cycle rate: increasing bending was observed in the belly of the leaflet with increasing frequency. The ratio of time of opening of the leaflets to time whilst fully open increased with the cycle rate, and the period of time when the valve was fully open and fully closed reduced markedly with increased cycle rate.

CONCLUSIONS

Leaflet kinematics, bending strains, the proportion of time the valve was fully closed or open, and the strain time integral were dependent on cycle rate in accelerated testing. This may influence the relative severity of accelerated testers.

Присъединете се към нашата
страница във facebook

Най-пълната база данни за лечебни билки, подкрепена от науката

  • Работи на 55 езика
  • Билкови лекове, подкрепени от науката
  • Разпознаване на билки по изображение
  • Интерактивна GPS карта - маркирайте билките на място (очаквайте скоро)
  • Прочетете научни публикации, свързани с вашето търсене
  • Търсете лечебни билки по техните ефекти
  • Организирайте вашите интереси и бъдете в крак с научните статии, клиничните изследвания и патентите

Въведете симптом или болест и прочетете за билките, които биха могли да помогнат, напишете билка и вижте болестите и симптомите, срещу които се използва.
* Цялата информация се базира на публикувани научни изследвания

Google Play badgeApp Store badge