BMC Pulmonary Medicine 2017-Jan

A novel reflex cough testing device.

কেবল নিবন্ধিত ব্যবহারকারীরা নিবন্ধগুলি অনুবাদ করতে পারবেন
প্রবেশ করুন - নিবন্ধন করুন
লিঙ্কটি ক্লিপবোর্ডে সংরক্ষিত হয়েছে
Kazunori Fujiwara
Katsuyuki Kawamoto
Yoko Shimizu
Takahiro Fukuhara
Satoshi Koyama
Hideyuki Kataoka
Hiroya Kitano
Hiromi Takeuchi

কীওয়ার্ডস

বিমূর্ত

The reflex cough test is useful for detecting silent aspiration, a risk factor for aspiration pneumonia. However, assessing the risk of aspiration pneumonia requires measuring not only the cough reflex but also cough strength. Currently, no reflex cough testing device is available that can directly measure reflex cough strength. We therefore developed a new testing device that can easily and simultaneously measure cough strength and the time until the cough reflex, and verified whether screening with this new instrument is feasible for evaluating the risk of aspiration pneumonia.

This device consists of a special pipe with a double lumen, a nebulizer, and an electronic spirometer. We used a solution of prescription-grade L-tartaric acid to initiate the cough reflex. The solution was inhaled through a mouthpiece as a microaerosol produced by an ultrasonic nebulizer. The peak cough flow (PCF) of the induced cough was measured with the spirometer. The 70 patients who participated in this study comprised 49 patients without a history of pneumonia (group A), 21 patients with a history of pneumonia (group B), and 10 healthy volunteers (control group).

With the novel device, PCF and time until cough reflex could be measured without adverse effects. The PCF values were 118.3 ± 64.0 L/min, 47.7 ± 38.5 L/min, and 254.9 ± 83.8 L/min in group A, group B, and the control group, respectively. The PCF of group B was significantly lower than that of group A and the control group (p < 0.0001), while that of group B was significantly lower than that of the control group (p < 0.0001). The time until the cough reflex was 4.2 ± 5.9 s, 7.0 ± 7.0 s, and 1 s in group A, group B, and the control group, respectively. This duration was significantly longer for groups A and B than for the control group (A: p < 0.001, B: p < 0.001), but there was no significant difference between groups A and B (p = 0.0907).

Our newly developed device can easily and simultaneously measure the time until the cough reflex and the strength of involuntary coughs for assessment of patients at risk of aspiration pneumonia.

আমাদের ফেসবুক
পেজে যোগদান করুন

বিজ্ঞানের দ্বারা সমর্থিত সবচেয়ে সম্পূর্ণ completeষধি ভেষজ ডেটাবেস

  • 55 ভাষায় কাজ করে
  • বিজ্ঞানের সহায়তায় ভেষজ নিরাময়
  • ইমেজ দ্বারা ভেষজ স্বীকৃতি
  • ইন্টারেক্টিভ জিপিএস মানচিত্র - অবস্থানের উপর গুল্ম ট্যাগ করুন (শীঘ্রই আসছে)
  • আপনার অনুসন্ধান সম্পর্কিত বৈজ্ঞানিক প্রকাশনা পড়ুন
  • তাদের প্রভাব দ্বারা herষধি গুল্মগুলি অনুসন্ধান করুন Search
  • আপনার আগ্রহগুলি সংগঠিত করুন এবং নিউজ রিসার্চ, ক্লিনিকাল ট্রায়াল এবং পেটেন্টগুলির সাথে আপ ডেট থাকুন

একটি লক্ষণ বা একটি রোগ টাইপ করুন এবং এমন গুল্মগুলি সম্পর্কে পড়ুন যা সহায়তা করতে পারে, একটি bষধি টাইপ করতে পারে এবং এর বিরুদ্ধে ব্যবহৃত রোগ এবং লক্ষণগুলি দেখতে পারে।
* সমস্ত তথ্য প্রকাশিত বৈজ্ঞানিক গবেষণার উপর ভিত্তি করে

Google Play badgeApp Store badge