Irish
Albanian
Arabic
Armenian
Azerbaijani
Belarusian
Bengali
Bosnian
Catalan
Czech
Danish
Deutsch
Dutch
English
Estonian
Finnish
Français
Greek
Haitian Creole
Hebrew
Hindi
Hungarian
Icelandic
Indonesian
Irish
Italian
Japanese
Korean
Latvian
Lithuanian
Macedonian
Mongolian
Norwegian
Persian
Polish
Portuguese
Romanian
Russian
Serbian
Slovak
Slovenian
Spanish
Swahili
Swedish
Turkish
Ukrainian
Vietnamese
Български
中文(简体)
中文(繁體)
Journal of Heart Valve Disease 2010-Jul

In-vivo assessment of a novel polymer (SIBS) trileaflet heart valve.

Ní féidir ach le húsáideoirí cláraithe ailt a aistriú
Logáil Isteach / Cláraigh
Sábháiltear an nasc chuig an gearrthaisce
Qiang Wang
Anthony J McGoron
Richard Bianco
Yasushi Kato
Leonard Pinchuk
Richard T Schoephoerster

Keywords

Coimriú

OBJECTIVE

A novel trileaflet polymer valve, which is a composite design of a biostable and biocompatible polymer poly(styrene-block-isobutylene-block-styrene) (SIBS) with an embedded reinforcement polyethylene terephthalate (PET) fabric, is being developed with the intention of providing a valve that has low thrombogenicity, high durability and favorable hemodynamic performance. The study aim was to investigate the biocompatibility and performance of this SIBS valve prototype under physiological loading conditions similar to humans, using a large-animal model.

METHODS

Four SIBS valves (two with surface modification using dimyristoyl phosphatidylcholine, DMPC), and two commercial Magna tissue valves, were implanted into sheep. Hemodynamic and blood chemistry measurements were performed periodically during the postoperative period. The explanted SIBS valves were extensively evaluated using macroscopic, histological, radiographical and scanning electron microscopy/energy-dispersive spectroscopy analysis.

RESULTS

Three animals, one with the DMPC-coated SIBS valve, and two with the Magna valves, reached the end of the study in satisfactory clinical condition, and were euthanized after 20 weeks. The other three animals (two with SIBS valves, one with a DMPC-coated SIBS valve) died at 6, 6.5, and 10 weeks due either to material failure or myocardial infarction. The explanted valves exhibited stent deformation and cracks on the leaflets, which exposed the underlying PET fabric and resulted in severe blood and tissue reactions. Extrinsic calcification was identified on the leaflets, and was associated with the regions of surface cracks.

CONCLUSIONS

The SIBS valve failed in animal testing because of material failure and calcification. The physical properties of SIBS must be improved in order to provide the structural integrity required for long-term in-vivo use in the form of a heart valve.

Bí ar ár
leathanach facebook

An bunachar luibheanna míochaine is iomláine le tacaíocht ón eolaíocht

  • Oibreacha i 55 teanga
  • Leigheasanna luibhe le tacaíocht ón eolaíocht
  • Aitheantas luibheanna de réir íomhá
  • Léarscáil GPS idirghníomhach - clibeáil luibheanna ar an láthair (ag teacht go luath)
  • Léigh foilseacháin eolaíochta a bhaineann le do chuardach
  • Cuardaigh luibheanna míochaine de réir a n-éifeachtaí
  • Eagraigh do chuid spéiseanna agus fanacht suas chun dáta leis an taighde nuachta, trialacha cliniciúla agus paitinní

Clóscríobh symptom nó galar agus léigh faoi luibheanna a d’fhéadfadh cabhrú, luibh a chlóscríobh agus galair agus comharthaí a úsáidtear ina choinne a fheiceáil.
* Tá an fhaisnéis uile bunaithe ar thaighde eolaíoch foilsithe

Google Play badgeApp Store badge