Nippon Hoshasen Gijutsu Gakkai zasshi 2003-Jul

[Evaluation of patient entrance dose rate of interventional X-ray equipment].

登録ユーザーのみが記事を翻訳できます
ログインサインアップ
リンクがクリップボードに保存されます
Shigeru Miyazaki
Shinji Abe
Yoh Katoh
Hiroyoshi Kobayashi
Hideo Uehara

キーワード

概要

Because interventional radiology (IVR) procedures are being performed with increasing frequency, patient X-ray exposure dose for X-ray fluoroscopic and radiographic procedures should not be ignored. In order to avoid excessive X-ray exposure, exposure dose rate limits are specified in the Japanese Industrial Standards (JIS) and by civil law at 50 mGy/min for usual fluoroscopy and 125 mGy/min for high-dose fluoroscopy. In the present study, we examined the difference in patient incident dose rate before and after using an X-ray generator that satisfied the above requirements. For incident dose to the image intensifier (I.I.), we investigated the differences between continuous and pulsed fluoroscopy, the effects of additional filters (Ta: tantalum, Al: aluminum), and the form of the X-ray spectrum. For pulsed fluoroscopy using PMMA (polymethyl-methacrylate), the maximum patient incident dose rates of usual and high-dose fluoroscopy were 59 mGy/min and 151 mGy/min, respectively. With regard to I.I. incident dose, saturation was observed beginning at a PMMA of 20 cm, and the X-ray dose was insufficient. In terms of the difference in patient incident dose rate with Ta and Al filters, the dose rate with the Ta filter was approximately 50% lower than that with the Al filter except for the saturation area. Concerning the X-ray spectrum, it was considered that a Ta filter not only minimizes patient X-ray exposure (because Ta reduces soft X-rays more effectively than Al) but also minimizes scattered X-rays because it filters out hard X-rays, leading to improved image quality. However, the use of the filter is appropriate only when a sufficient I.I. incident dose can be ensured. Specifically, the use of the filter under saturation conditions can lead to deterioration in image quality. Therefore, IVR X-ray systems must be equipped with an appropriate filter for reducing X-ray exposure while maintaining a sufficient I.I. incident dose rate.

Facebookページに参加する

科学に裏打ちされた最も完全な薬草データベース

  • 55の言語で動作します
  • 科学に裏打ちされたハーブ療法
  • 画像によるハーブの認識
  • インタラクティブGPSマップ-場所にハーブをタグ付け(近日公開)
  • 検索に関連する科学出版物を読む
  • それらの効果によって薬草を検索する
  • あなたの興味を整理し、ニュース研究、臨床試験、特許について最新情報を入手してください

症状や病気を入力し、役立つ可能性のあるハーブについて読み、ハーブを入力して、それが使用されている病気や症状を確認します。
*すべての情報は公開された科学的研究に基づいています

Google Play badgeApp Store badge