Japanese
Albanian
Arabic
Armenian
Azerbaijani
Belarusian
Bengali
Bosnian
Catalan
Czech
Danish
Deutsch
Dutch
English
Estonian
Finnish
Français
Greek
Haitian Creole
Hebrew
Hindi
Hungarian
Icelandic
Indonesian
Irish
Italian
Japanese
Korean
Latvian
Lithuanian
Macedonian
Mongolian
Norwegian
Persian
Polish
Portuguese
Romanian
Russian
Serbian
Slovak
Slovenian
Spanish
Swahili
Swedish
Turkish
Ukrainian
Vietnamese
Български
中文(简体)
中文(繁體)
Annals of Thoracic Surgery 1995-Aug

New evolution in mitral physiology and surgery: mitral stentless pericardial valve.

登録ユーザーのみが記事を翻訳できます
ログインサインアップ
リンクがクリップボードに保存されます
R F Deac
D Simionescu
D Deac

キーワード

概要

The human adult mitral valve, with a mean diastolic area of up to 7.6 cm2, excess leaflet surface area for coaptation in systole, mitral annulus-papillary muscle continuity, and systolic constriction of the posterior left ventricular wall around the mitral annulus functions in concert with other components of the left side of the heart. Mitral valve replacement with an artificial valve that interferes with the normal physiology could account for less than adequate late results. A stentless biologic mitral valve substitute has been designed, constructed, and tested. The size of the valve is selected according to the circumference of the excised valve within certain limits. The valve is manufactured of two square or trapezoidal pieces of selected stabilized human autologous or bovine pericardium. The pericardial pieces are sutured together by their lateral margins, thus creating a frusto-conical valvular body. The upper circumference of the valvular body is sutured at the mitral annulus and the lower margin with the new chordae is attached by suture at each papillary muscle. In vitro testing of six stentless bovine pericardial valves in a Rowan-Ash fatigue tester at 1,200 cycles/min revealed a durability of more than 320 million cycles. Clinical use of described technique initiated in 1989 was performed in 18 patients by one surgeon (30 patients in the same institution). The mean valve size was 29 mm circularized diameter. There was no mortality in this group of patients up to 70 months of follow-up. Valve competence was obtained in every case by adequate sizing of the valve.(ABSTRACT TRUNCATED AT 250 WORDS)

Facebookページに参加する

科学に裏打ちされた最も完全な薬草データベース

  • 55の言語で動作します
  • 科学に裏打ちされたハーブ療法
  • 画像によるハーブの認識
  • インタラクティブGPSマップ-場所にハーブをタグ付け(近日公開)
  • 検索に関連する科学出版物を読む
  • それらの効果によって薬草を検索する
  • あなたの興味を整理し、ニュース研究、臨床試験、特許について最新情報を入手してください

症状や病気を入力し、役立つ可能性のあるハーブについて読み、ハーブを入力して、それが使用されている病気や症状を確認します。
*すべての情報は公開された科学的研究に基づいています

Google Play badgeApp Store badge