American Journal of Critical Care 2014-Sep

Cleanliness of disposable vs nondisposable electrocardiography lead wires in children.

Само регистрираните корисници можат да преведуваат статии
Пријавете се / пријавете се
Врската е зачувана во таблата со исечоци
Nancy Addison
Beth Quatrara
Lisa Letzkus
David Strider
Virginia Rovnyak
Virginia Syptak
Lisa Fuzy

Клучни зборови

Апстракт

BACKGROUND

Mediastinitis costs hospitals thousands of dollars a year and increases the incidence of patient morbidity and mortality. No studies have been done to evaluate adenosine triphosphate (ATP) counts on disposable and nondisposable electrocardiography (ECG) lead wires in pediatric patients.

OBJECTIVE

To compare the cleanliness of disposable and nondisposable ECG lead wires in postoperative pediatric cardiac surgery patients by measuring the quantity of ATP (in relative luminescence units [RLUs]). ATP levels correlate with microbial cell counts and are used by institutions to assess hospital equipment and cleanliness.

METHODS

A prospective, randomized trial was initiated with approval from the institutional review board. Verbal consent was obtained from the parents/guardians for each patient. Trained nurses performed ATP swabs on the right and left upper ECG cables on postoperative days 1, 2, and 3.

RESULTS

This study enrolled 51 patients. The disposable ECG lead wire ATP count on postoperative day 1 (median, 157 RLUs) was significantly lower (P < .001) than the count for nondisposable ATP lead wires (median, 610 RLUs). On postoperative day 2, the ATP count for the disposable ECG lead wires (median, 200 RLUs) was also lower (P = .06) than the count for the nondisposable ECG lead wires (median, 453 RLUs).

CONCLUSIONS

Results of this study support the use of disposable ECG lead wires in postoperative pediatric cardiac surgery patients for at least the first 48 hours as a direct strategy to reduce the ATP counts on ECG lead wires.

Придружете се на нашата
страница на Facebook

Најкомплетната база на податоци за лековити билки поддржана од науката

  • Работи на 55 јазици
  • Лекови од билки поддржани од науката
  • Препознавање на билки по слика
  • Интерактивна GPS мапа - означете ги билките на локацијата (наскоро)
  • Прочитајте научни публикации поврзани со вашето пребарување
  • Пребарувајте лековити билки според нивните ефекти
  • Организирајте ги вашите интереси и останете во тек со истражувањето на новостите, клиничките испитувања и патентите

Напишете симптом или болест и прочитајте за билки што можат да помогнат, напишете билка и видете болести и симптоми против кои се користи.
* Сите информации се базираат на објавени научни истражувања

Google Play badgeApp Store badge