Investigative Ophthalmology and Visual Science 2014-Apr

Retinal oximetry with a scanning laser ophthalmoscope.

Samo registrirani uporabniki lahko prevajajo članke
Prijava / prijava
Povezava se shrani v odložišče
Jona Valgerdur Kristjansdottir
Sveinn Hakon Hardarson
Gisli Hreinn Halldorsson
Robert A Karlsson
Thorunn S Eliasdottir
Einar Stefánsson

Ključne besede

Povzetek

OBJECTIVE

The purpose of the study was to assess if a scanning laser ophthalmoscope (SLO), Optomap 200Tx, could be used for measurements of hemoglobin oxygen saturation in retinal blood vessels.

METHODS

Optomap 200Tx uses two lasers for image acquisition, 532 and 633 nm. Retinal images of healthy individuals and patients with retinal vein occlusion were analyzed with modified Oxymap Analyzer software, which tracks retinal vessels and calculates relative hemoglobin oxygen saturation.

RESULTS

Oxygen saturation in healthy individuals was measured as 92% ± 13% for arterioles and 57% ± 12% for venules (mean ± SD, n = 11, P = 0.0001). Standard deviation for repeated measurements of the same eye was 3.5% for arterioles and 4.4% for venules. In patients with confirmed venular hypoxia, central retinal vein occlusion (CRVO) or hemivein occlusion, the average venular oxygen saturation was measured as 23% ± 3% in the affected eyes and 59% ± 3% in the fellow eyes (n = 4, P = 0.0009).

CONCLUSIONS

Technically, it is possible to derive information on retinal oxygen saturation from an SLO with a 2-wavelength oximetry algorithm. The system produced both sensitive and repeatable results. The remaining challenges include decreasing variability between vessels of the same eye and variability between individuals. Given the advantages that SLO imaging has over conventional fundus camera optics in retinal oximetry, further development of SLO oximetry may provide the optimal approach to retinal oximetry.

Pridružite se naši
facebook strani

Najbolj popolna baza zdravilnih zelišč, podprta z znanostjo

  • Deluje v 55 jezikih
  • Zeliščna zdravila, podprta z znanostjo
  • Prepoznavanje zelišč po sliki
  • Interaktivni GPS zemljevid - označite zelišča na lokaciji (kmalu)
  • Preberite znanstvene publikacije, povezane z vašim iskanjem
  • Iščite zdravilna zelišča po njihovih učinkih
  • Organizirajte svoje interese in bodite na tekočem z raziskavami novic, kliničnimi preskušanji in patenti

Vnesite simptom ali bolezen in preberite o zeliščih, ki bi lahko pomagala, vnesite zelišče in si oglejte bolezni in simptome, proti katerim se uporablja.
* Vse informacije temeljijo na objavljenih znanstvenih raziskavah

Google Play badgeApp Store badge