Prizma Pulse Oximeter Evaluation
Кључне речи
Апстрактан
Опис
The device under test was the Prizma which is a phone cover biometric device that includes reflectance oximetry. This device is a spot checker that requires the patient to hold their finger on the sensor during the measurement. The subjects were asked not to move during the study. In order to get consistent measurements the fingers were taped to the test device. The tape acted mainly as a reminder, and there were some problems with finger movement. A radial arterial cannula was placed in either the left or right wrist of each subject.
Blood gas analysis to determine oxyhemoglobin saturation was performed on an OSM-3® multi-wavelength oximeter (Hemoximeter, Radiometer, Copenhagen). No subject was anemic (Hemoglobin ≤ 10gm•dl-1). Each subject had control data taken at the beginning of each experiment, with two control blood samples drawn while breathing room air. Hypoxia was induced to different levels of oxyhemoglobin saturation (between 70-100%) by having subjects breathe mixtures of nitrogen, room air, and carbon dioxide. Oxyhemoglobin saturation was reduced to a series of targets and stabilized at the plateau value. Each plateau level of oxyhemoglobin saturation was maintained for at least 30 seconds. Two arterial blood samples were then obtained, approximately 30 seconds apart. A total of 24 samples were obtained per subject. Data were recorded by Bickler-Ye lab and provided for analysis.
Датуми
Последња верификација: | 02/28/2017 |
Фирст Субмиттед: | 02/26/2017 |
Предвиђена пријава послата: | 03/08/2017 |
Прво објављено: | 03/14/2017 |
Послато последње ажурирање: | 03/08/2017 |
Последње ажурирање објављено: | 03/14/2017 |
Стварни датум почетка студије: | 04/14/2016 |
Процењени датум примарног завршетка: | 04/14/2016 |
Предвиђени датум завршетка студије: | 04/15/2016 |
Стање или болест
Интервенција / лечење
Device: Prizma
Фаза
Групе руку
Арм | Интервенција / лечење |
---|---|
Prizma Measurement of oxyhemoglobin saturation by OSM-3 oximeter and Prizma oxygen saturation measurement | Device: Prizma A radial arterial cannula was placed in either the left or right wrist of each subject.
Blood gas analysis to determine oxyhemoglobin saturation was performed on an OSM-3® multi-wavelength oximeter (Hemoximeter, Radiometer, Copenhagen). No subject was anemic (Hemoglobin ≤ 10gm•dl-1). |
Критеријуми
Узраст подобан за студирање | 21 Years До 21 Years |
Полови подобни за студирање | All |
Метода узорковања | Non-Probability Sample |
Прихвата здраве волонтере | да |
Критеријуми | Inclusion Criteria: - Both male and female subjects - Healthy subjects capable of undergoing controlled hypoxemia to the levels outlined in the desaturation profile below. - Meeting the demographic requirements listed above. Exclusion Criteria: - Smokers or individuals exposed to high levels of carbon monoxide that result in carboxyhemoglobin levels over 3% as tested by the BYL. - Subjects who would be placed under medical risk under the desaturation profile listed below. - Significant arrhythmia - Age below 21 or over 50 - Pregnant women |
Исход
Примарне мере исхода
1. Blood oxygen saturation [2 days]