An automated system for the measurement of alanine/EPR dosimeters
關鍵詞
抽象
NPL for several years has offered mailed reference dosimetry services based on alanine/EPR dosimeters, both at industrial and therapy dose levels. Compared to other methods of reference dosimetry, operator involvement in alanine/EPR has been found to be relatively high, and contributes significantly to the overall economics of the process. Commercially available sample changers are not suitable for high accuracy applications, and it has proved necessary to develop a dedicated automation system to handle NPL alanine dosimeter pellets. In this paper we describe an automatic sample changer for placing and retrieving alanine pellets into and out of the cavity of a standard research grade EPR spectrometer. Up to 32 pellets can be held in each removable sample tray. The sample changer software has been interfaced into the spectrometer control software to enable complete automation of the measurement process, including the optimization of spectrometer settings and rotation of the sample within the cavity.